Download Catalog for Benchtop Conductivity Meter LBCM-A14

Benchtop Conductivity Meter LBCM-A14

Benchtop Conductivity Meter LBCM-A14

www.labtron.co.uk | info@labtron.co.uk

Overview

Benchtop Conductivity Meter LBCM-A14 with 1 to 3 points push button conductivity calibration, automatically recognizes calibration standards. With setup menu, it does setting of 6 parameters calibration points, temperature units and others. Large LCD display shows settings and measurement values. Temperature calibration by manual method improves the accuracy of conductivity meter.

Specifications :

Conductivity Mode <table class='table table-striped table-bordered table-responsive'><tr><td><strong>Conductivity measurement range</strong></td><td>0.01-20.0, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm</td></tr><tr><td><strong>Conductivity accuracy</strong></td><td>±1% F.S.</td></tr><tr><td><strong>Conductivity resolution</strong></td><td>0.001, 0.01, 0.1, 1</td></tr><tr><td><strong>Conductivity calibration points</strong></td><td>1 to 3 points</td></tr><tr><td><strong>Conductivity calibration solutions</strong></td><td>10 µS/cm , 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm</td></tr></table>
TDS Mode <table class='table table-striped table-bordered table-responsive'><tr><td><strong>TDS measurement Range</strong></td><td>0 to 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)</td></tr><tr><td><strong>TDS Accuracy</strong></td><td>± 1 % F.S.</td></tr><tr><td><strong>TDS Factor</strong></td><td>0.01 to 1.00 (default 0.5)</td></tr><tr><td><strong>TDS Calibration points</strong></td><td>3 points</td></tr></table>
Temperature Mode <table class='table table-striped table-bordered table-responsive'><tr><td><strong>Temperature range</strong></td><td>0 to 105°C, 32 to 221°F</td></tr><tr><td><strong>Temperature accuracy</strong></td><td>±1°C</td></tr><tr><td><strong>Temperature resolution</strong></td><td>0.1°C, 0.1°F</td></tr><tr><td><strong>Temperature calibration points</strong></td><td>1 point</td></tr><tr><td><strong>Temperature Compensation Range and Mode</strong></td><td>Manual or Automatic, 0 to 100°C , 32 to 212°F</td></tr><tr><td><strong>Temperature coefficient</strong></td><td>0.0 to 10.0 % per °C</td></tr></table>
Cell constant K=0.1, 1, 10
Power requirements DC 9 V, using AC adapter, AC 220 V, 50 Hz
Dimension (L×W×H) 210 × 205 × 75 mm
Weight 1.5 kg

Features :

  • -  Measures conductivity, TDS, Temperature of the liquids
  • -  Automatic temperature compensation for accurate values over the entire range
  • -  Selectable temperature units (°C or °F) for different application requirements
  • -  Reset function automatically resumes all the setting back to factory default options
  • -  Manual and automatic mode of temperature calibration
  • -  Auto hold feature to lock, record and read stable measurement endpoint
  • -  Stability indicator displays the status of current measurement
  • -  Help message as an operational guide, with quick access to test parameters
  • -  Conductivity/TDS Mode
  • -  1 to 3 Conductivity push button calibration points
  • -  3 TDS calibration points
  • -  1 Temperature calibration point